Control unit of the Integral XT -
Easy operation via the Command console

The process thermostats are operated via the Command console which is also used in the LAUDA Proline range. The console is easily removable and can be used as a remote control. The easy menu-driven operation makes operation intuitive and simplifies fast changes of the settings. An additional extension cable on the Command console enables operation of the device and process surveillance from a distance of up to 50 metres. An opto-decoupled RS 232/485 interface is integrated in the Command console as standard for connection to a PC.

Performance profile of the Integral XT:

- High-resolution, back-lit graphic LC display for depicting the entire
  thermostating system, and with individually-adjustable
  surface
- Resolution of the actual value reading optional up to 0.001°C
- Real-time clock combined with timer function
- Editable programmer for up to 150 segments in 5 programs
- Nine pre-defined sets of parameters can be selected for the control
- Menu-driven operation in German, English, French or Spanish
- Function lights for heating, cooling and external control
- Error reports by means of bright LED lights, acoustic alarm and clear
  text message
- Adjustable limit of the pump pressure
- Integrated external control with cascade connection
- Supply temperature limit for the gentle heating of the external system
- Integrated opto-decoupled RS 232/484 interface as standard

Anwendung1 Kontrollkopf

The entire process at a glance

The clear process diagram on the high-resolution graphic display of the Command console displays all the device parameters necessary for the thermostating process at a glance. Alternatively, the user can define other views and, hence, can always see the information which is important for him.

Vorteile6

Flexible interface concept

A multitude of individual options is available for connecting peripheral devices. The units are equipped with a Pt 100 temperature probe connection and two sockets for components with the LAUDA internal LiBus such as the Command control unit, as standard. Similar to Proline, optional modules for fitting up to two plug-in modules are available. The following modules are currently available:

RS 232/485 Interface - Cat. No. LRZ 013
With opto coupler galvanically isolated, 9-pole SUB-D socket. With
LAUDA instruction set as far as possible compatible with LAUDA
Ecoline and LAUDA Integral T series.


Analogue module - Cat. No. LRZ 912
With two inputs and two outputs on 6-pole DIN socket. The
inputs and outputs are independent of one another, adjustable
as 4...20 mA, 0...20 mA or 0...10 V interface.

- The following input values can be predefined: set temperature,
  external actual temperature, pump output.
- Adjusting the output values: set temperature, actual temperature
  (bath temperature), external actual temperature from Pt
  100, external actual temperature from analogue input, correcting
  variable, pump output, pump rpm. Apart from this, the interfaces
  can be freely adjusted in accordance with the set function.
  For instance: 0 mA equals 0 °C and 20 mA equals 100 °C.


Contact module SUB-D socket - Cat. No. LRZ 915
15-pole SUB-D socket with three relay contact-outputs (centre
zero relay, maximum 30V/0, 2A) and three binary inputs for controlling
via external potential free contacts.

- Functions of the inputs: Setting of faults or standby, control of
  the programmer, half duplex operation (two different set
  temperatures) or control type (internal external control).
- Functions of the outputs: Signalling the status of diverse faults,
  standby, or top up bath medium, output the window discriminator
  status (inside outside) or the programmer status.
  Functions scheme contact module SUB-D.


Contact module NAMUR - Cat. No. LRZ 914
Contact module with plug and socket in accordance with NAMUR
NE28 standard, functionality as for above, but only one input and
one output on two DIN sockets.


Profibus interface - Cat. No. LRZ 917
For inclusion of up to 120 devices into a Profibus network
Connection with 9-pin SUB D-socket, transfer rates up to 12Mbaud, integrated LED for the display of connection errors.